Appendix B. Status Table and Settings
531
Table 124. Status-Table Fields and Descriptions
Fieldname Description Variable Type Default Range Edit? Info Type
RevBoard
xxx.yyy
xxx = hardware revision
number; yyy = clock chip
software revision; stored in
FLASH memory.
Integer Status
StationName1
Sets a name internal to the
CR3000. Stored in flash
memory. Not to be confused
with the station name set in
datalogger support software.
See foot note for limitations.
String Yes Config
PakBusAddress2
CR3000 PakBus address. String 1 1 to 3999 Yes Config PB
ProgName
Name of current (running)
program.
String Status
StartTime
Time the program began
running.
Time Status
RunSignature
Signature of the compiled
binary data structure for the
current program. Value is
independent of comments
added or non-functional
changes to the program.
Often changes with
operating-system changes.
Integer Status
ProgSignature
Signature of the current
running program file
including comments. Does
not change with operating
system changes.
Integer Status
Battery
Current value of the battery
voltage. Measurement is
made in the background
calibration.
Float 9.6-16 Vdc Measure
PanelTemp
Current wiring-panel
temperature. Measurement is
made in the background
calibration.
Float Measure
WatchdogErrors3
Number of Watchdog errors
that have occurred while
running this program.
Integer 0 0
Yes
Reset by
changing to 0
Error
LithiumBattery4
Current voltage of the
lithium battery. Measurement
is updated in background
calibration.
Float 2.7-3.6 Vdc Measure
Komentáře k této Příručce