Campbell-scientific CR3000 Micrologger Uživatelský manuál Strana 473

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 590
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 472
Section 11. Glossary
473
watchdogtimer
Anerrorcheckingsystemthatexaminestheprocessorstate,software
timers,andprogramrelatedcounterswhenthedataloggerisrunningits
program.Iftheprocessorhasbombedorisneglectingstandardsystem
updatesorifthecountersareoutsidethelimits,thewatchdogtimer
resetstheprocessorandprogramexecutio
n.Voltagesurgesand
transientscancausethewatchdogtimertoresettheprocessorand
programexecution.Whenthewatchdogtimerresetstheprocessorand
programexecution,anerrorcountisincrementedinthe
WatchdogTimerentryoftheStatustable
(p.530).Alownumber(1to10)
ofwatchdogtimerresetsisofconcern,butnormallyindicatestheuser
shouldjustmonitorthesituation.Alargenumber(>10)oferrors
accumulatingoverashortperiodoftimeshouldcauseincreasingalarm
sinceitindicatesahardwareorsoftwareproblemmayexist.Whenlar
ge
numbersofwatchdogtimerresetsoccur,consultwithaCampbell
Scientificapplicationsengineer.
weathertight
Describesaninstrumentationenclosureimpenetrablebycommon
environmentalconditions.Duringextraordinaryweatherevents,
however,sealsontheenclosuremaybebreached.
WebAPI
ApplicationProgrammingInterface(seesectionWebAPI
,formore
information).
Glossary.WildCard
acharacterorexpressionthatsubstitutesforanyothercharacteror
expression.
XML
Extensiblemarkuplanguage.
UserProgram
TheCRBasicprogramwrittenbytheCR3000userinCRBasicEditoror
ShortCut.
11.2 Concepts
11.2.1 Accuracy, Precision, and Resolution
Three terms often confused are accuracy, precision, and resolution. Accuracy is a
measure of the correctness of a single measurement, or the group of
measurements in the aggregate. Precision is a measure of the repeatability of a
Zobrazit stránku 472
1 2 ... 468 469 470 471 472 473 474 475 476 477 478 ... 589 590

Komentáře k této Příručce

Žádné komentáře