CS110 Electric Field MeterRevision: 4/12Copyright © 2005-2012Campbell Scientific, Inc.

Strany 2

CS110 Electric Field Meter Unlike traditional rotating vane field mills, the CS110 uses a reciprocating shutter. A stepper motor opens and then close

Strany 3 - Warranty

CS110 Electric Field Meter 1.5 Digital I/O Three general purpose 0 to 5 V digital I/O lines are available on the CS110 Power cable (CS110CBL3-L) that

Strany 4 - Assistance

CS110 Electric Field Meter 2. CS110 Specifications Electric Field Measurement Performance: Parallel-Plate Configuration Accuracy ±1% of reading + 60

Strany 5 - CS110 Table of Contents

CS110 Electric Field Meter Sample (Measurement) Rate: Programmable sample rate up to 5 samples per second, variable sample rates possible. Variable e

Strany 6

CS110 Electric Field Meter Rugged Construction: Ultra-reliable metallic ground connection to reciprocating shutter (no wiping contact), brushless ste

Strany 7 - Figures

CS110 Electric Field Meter 3. CS110 Measurement Details The charge amplifier circuitry of the reciprocating electric field meter is depicted in Figur

Strany 8

CS110 Electric Field Meter 160 FIGURE 3. Charge amplifier output during an electric field measurement cycle Offset voltages Voff1 and Voff2 are zero

Strany 9

CS110 Electric Field Meter Ileak = Cf·(Voff1 – Voff2)/ΔT + Icomp (eq. 2) Where Cf is the value of feedback capacitor used in the charge amplifier,

Strany 10 - 1.3 Meteorological Inputs

CS110 Electric Field Meter 11010010000.1 1 10 100Measurement Interval (Seconds)Average Current (mA) @ 12 V FIGURE 4. CS110 average current consumptio

Strany 11 - 1.6 Self-Check Features

CS110 Electric Field Meter CS110 CS I/O cable (CS110CBL2-L). Examples of CS I/O peripherals include the CR1000 Keyboard Display and the COM220 phone

Strany 13

CS110 Electric Field Meter minimize non-ideal fringing effects. Sharp corners were avoided in order to prevent corona discharge. All metal parts of

Strany 14

CS110 Electric Field Meter of variations in surface cleanliness along with charging and discharging processes. Polished 316-L stainless-steel is used

Strany 15

CS110 Electric Field Meter FIGURE 7. NIST calibration certificate Careful removal and replacement of the stator on the CS110 does not invalidate the

Strany 16 - potentials

CS110 Electric Field Meter Consequently, a downward facing and elevated configuration as illustrated in Figure 8 is recommended for long-term field ap

Strany 17 - 4.1 Power Requirements

CS110 Electric Field Meter FIGURE 9. Campbell Scientific, Inc. electric field meter site correction facility 16

Strany 18 - 4.3 Communication Options

CS110 Electric Field Meter An upward-facing calibration kit (PN: 17579) was developed to hold the CS110 in a flush-mounted upward-facing position, as

Strany 19 - 5.1 Factory Calibration

CS110 Electric Field Meter utilized to autonomously measure, process and store data to aid in site correction. Campbell Scientific, Inc. has performe

Strany 20

CS110 Electric Field Meter covering the stator with a clean Zero Electric Field Cover (PN: 17642). Fair weather field site correction is not recommen

Strany 21 - <⏐30 V/m⏐, which

CS110 Electric Field Meter August 2, 2005 Thunderstorm at Logan, Utah-5000-4000-3000-2000-100001000200030004000500060007000800090005:34:00 5:40:00 5:4

Strany 22 - 5.2 Site Correction

CS110 Electric Field Meter A network or array of electric field meters improves lightning warning because of a wider area of coverage along with measu

Strany 23

Warranty “PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materi

Strany 24

CS110 Electric Field Meter Public variables are defined and available for viewing in the Public table, which is a data table automatically set up by t

Strany 25

CS110 Electric Field Meter Units wind_speed=mph Public wind_dir Units wind_dir=deg Public solar_rad Units solar_rad=W/m2 Public air_temp Units air_tem

Strany 26

CS110 Electric Field Meter Scan(1,sec,0,0) for k = 1 to 16 'Initialize status array. E_status(k) = 0 next PanelTemp (panel_temp,250) Bat

Strany 27 - 6. Lightning Warning

CS110 Electric Field Meter accomplished using a disable variable (DisableVar) associated with appropriate output processing instructions. The last pa

Strany 28

CS110 Electric Field Meter Single-ended voltage measurements are referenced to ground, rather than the low side of a differential input. The VoltSE si

Strany 29 - 7. CRBasic Programming

CS110 Electric Field Meter determine if results are returned as counts for a given interval (POption = 0), or as frequency = counts/(scan interval in

Strany 30

CS110 Electric Field Meter purchased and connected to the SOLAR RADIATION connector to interface to either the CS100 or the CS106 barometric pressure

Strany 31

CS110 Electric Field Meter 8.7 Measuring Rainfall Circular connector labeled RAIN can be used to connect up a rain gauge using a switch closure such

Strany 32

CS110 Electric Field Meter default PakBus address of 1. Use LoggerNet Tool Bar’s “Setup” button or “EZ Setup” button to create the PC to CS110 commun

Strany 33 - provides for the

CS110 Electric Field Meter Remember to click on the “Apply” button to cause the settings to take effect. The “Apply” button is grayed out once it ha

Strany 34

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Strany 35

CS110 Electric Field Meter If the connection is made and the stations time shows up in the window, you can then select the “Numeric:” button and add t

Strany 36

CS110 Electric Field Meter 10. Maintenance 10.1 Checking Site Ground Integrity The CS110 electric field meter needs to be electrically connected to

Strany 37 - 9. PC Software

CS110 Electric Field Meter power electronics to high voltage switchgear. NO-OX-ID A-SPECIAL prevents the formation of oxides, sulfides and other corr

Strany 38

CS110 Electric Field Meter • Turbine studs • Coal crushers • Casting and Molding equipment • Heat exchangers 10.3 Self-Check Features The CS1

Strany 39

CS110 Electric Field Meter 10.4 Cleaning the CS110 Electrode Head The CS110 motor assembly illustrating the 316-L stainless-steel stator, shutter, an

Strany 40

CS110 Electric Field Meter ±4.2 nA. Leakage current values in excess of ±4.2 nA can cause measurement errors and are indicated by status = 11. Cleani

Strany 41 - 10. Maintenance

CS110 Electric Field Meter VoltDiff(Internal_RH,1,mV2500,5,True,0,250,0.1,0) 'Internal humidity measure. Changing of CS110 desiccant is recommend

Strany 42

CS110 Electric Field Meter 2. Inspect the gasket on the CS110 lid making sure that a good seal is possible when the lid is replaced. 3. Remove the

Strany 43 - 10.3 Self-Check Features

CS110 Electric Field Meter 40 10.7 Re-Calibration Re-calibration of measurement instruments is commonly done in data critical applications in order t

Strany 44

Appendix A. CS110 Measurement Status Codes Status codes 1 through 3: Good Instrument Health. status = 1, Good instrument health. ±250 mV measureme

Strany 45 - 10.5 Changing Desiccant

CS110 Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to

Strany 46

Appendix A. CS110 Measurement Status Codes A-2 Status codes 11 through 16: Measurement warnings and errors. status = 11, Leakage current exceeds co

Strany 47

Appendix B. CS110 Accessories B.1 Zero Electric Field Cover As previously mentioned, unwanted surface charges residing on non-conductive deposits on

Strany 48 - 11. References

Appendix B. CS110 Accessories B-2

Strany 49 - Status Codes

Appendix C. CS110 Connector Pin-outs The following information describes the connectors that mate with the built-in (bulkhead) connectors on the CS11

Strany 50 - > ⏐1.00 V⏐. Return

Appendix C. CS110 Connector Pin-outs RAIN CONNECTOR Connector is male and is CSI PN 9889 (Conxall #6282-6PG-522 or Switchcraft #EN3C6M). Pin 1 is in

Strany 51 - B.3 CR1000 Keyboard Display

Appendix C. CS110 Connector Pin-outs SOLAR RADIATION CONNECTOR Connector is male and is CSI PN 9889 (Conxall #6282-6PG-522 or Switchcraft #EN3C6M).

Strany 52

Appendix C. CS110 Connector Pin-outs C-4

Strany 53

Appendix D. Servicing the CS110 The CS110 has been designed to provide reliable electric field measurements and to simplify and minimize maintenance.

Strany 54

Appendix D. Servicing the CS110 CS110 panel board, after removing the two thumb screws on the CR1000 module. When installing a CR1000 into the CS110

Strany 55

Appendix D. Servicing the CS110 D.5 Shutter/Encoder Alignment A factory trim is done to align the position of the CS110 shutter with an Index mark o

Strany 56

CS110 Table of Contents 10. Maintenance ...33 10.1 Checking Site Ground Integrity ...

Strany 57 - D.2 Changing Out the CR1000

Appendix D. Servicing the CS110 EndIf If Flag(2) Then CS110Trim(-1) 'Take a step closed. Flag(2) = 0 'Reset Flag 2. EndIf NextSca

Strany 58

Appendix D. Servicing the CS110 Motion of Encoder Base PlateEncodePlate Connector Locking Plate r Base FIGURE D-2. CS110 motor assembly Referring t

Strany 59

Appendix D. Servicing the CS110 'Program to open/close the CS110 shutter (CS110_Shutter1.cr1). 'Last updated by Jody Swenson on 9/26/05.

Strany 60

Appendix E. CS110 as a Slow Antenna As previously mentioned the CS110 can sample the external electric field at a maximum rate of 5 Hz (200 ms) using

Strany 61

Appendix E. CS110 as a Slow Antenna 0.1 1 10 100 1103× 1104×00.1670.3330.50.6670.8331Overall Frequency Response of CS110 Slow-Antenna Measurement wit

Strany 62 - D.6 Motor O-ring Seal

Appendix E. CS110 as a Slow Antenna E.2 Response of the CS110 Slow Antenna in the Time Domain The following graphs shows one lightning strike measur

Strany 63 - trueof 0.708==−

Appendix E. CS110 as a Slow Antenna FIGURE E-3. KSC electric field change and CS110 slow antenna data The KSC electric field mill and the CS110 wer

Strany 64

Appendix E. CS110 as a Slow Antenna E.3 Programming The following CRBasic program utilizes the slow antenna capability of the CS110. 'Program

Strany 65

Appendix E. CS110 as a Slow Antenna E-6 E.4 Calibration The factory calibration described in Section 5 applies to the maximum amplitude of the step

Strany 66

Appendix F. Example CRBasic Programs An example CS110 weather station program using a variable electric field measurement rate in order to minimize c

Strany 67 - E.3 Programming

CS110 Table of Contents H. Tripod CS110 and StrikeGuard Site... H-1 H.1 Tripod CS110 and StrikeGuard ...

Strany 68 - E.4 Calibration

Appendix F. Example CRBasic Programs DataTable(Tabslow,1,-1) '-1 to auto-allocate all available memory. DataInterval(0,60,sec,10) 'Avera

Strany 69 - Programs

Appendix F. Example CRBasic Programs CS110(E_field,leakage_cur,status,_60Hz,Mcorrected,0) If E_field = NAN Then meas_error = 1 'Disa

Strany 70

Appendix F. Example CRBasic Programs F-4

Strany 71

Appendix G. CS110 2 Meter CM10 Tripod Site FIGURE G-1. CS110 2 Meter CM10 Tripod site The Tripod_CS110 site consists of the CS110 mounted on the CM

Strany 72

Appendix G. CS110 2 Meter CM10 Tripod Site Installation of the Tripod_CS110 Site: Drive the ground rod into the ground where the center of the tripod

Strany 73 - Tripod Site

Appendix G. CS110 2 Meter CM10 Tripod Site FIGURE G-2. CS110 on CM10 Tripod Mast Mount the fiberglass enclosure to the top of the leg facing the eq

Strany 74

Appendix G. CS110 2 Meter CM10 Tripod Site FIGURE G-3. Earth grounding Stake the tripod to the ground and/or weigh it down with sand bags. Determin

Strany 75

Appendix G. CS110 2 Meter CM10 Tripod Site Site Correction for Tripod CS110 Only Site - October 2, 20051 minute average of 1 second dataResults indic

Strany 76

Appendix G. CS110 2 Meter CM10 Tripod Site G-6

Strany 77

Appendix H. Tripod CS110 and StrikeGuard Site H.1 Tripod CS110 and StrikeGuard FIGURE H-1. Tripod CS110 and StrikeGuard The Tripod CS110 and Strik

Strany 78

CS110 Table of Contents iv

Strany 79 - StrikeGuard Site

Appendix H. Tripod CS110 and StrikeGuard Site H.1.1 Installation of the Tripod CS110 and StrikeGuard Site Drive the ground rod into the ground where

Strany 80

Appendix H. Tripod CS110 and StrikeGuard Site FIGURE H-2. CS110 and StrikeGuard on tripod mast Mount the fiberglass enclosure to the top of the leg

Strany 81

Appendix H. Tripod CS110 and StrikeGuard Site FIGURE H-3. Grounding the CS110 grounding strap H-4

Strany 82

Appendix H. Tripod CS110 and StrikeGuard Site FIGURE H-4. Grounding the tripod and battery Stake the tripod to the ground or weight it down with san

Strany 83

Appendix H. Tripod CS110 and StrikeGuard Site FIGURE H-5. Connections for combined system H-6

Strany 84

Appendix H. Tripod CS110 and StrikeGuard Site Power and communication cables should be run down the equator side of the mast and under the enclosure

Strany 85

Appendix H. Tripod CS110 and StrikeGuard Site H-8

Strany 87

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel

Strany 88

CS110 Electric Field Meter 1. General Description 1.1 CS110 Introduction Atmospheric electric fields have been measured for decades by electric fiel