Section 8. Operation
308
Error Calculations with Input Reversal = True
µVerror=gainterm+offsetterm
=(830.7µV*0.12%)+(1.5*0.67µV+1.0µV)
=0.997µV+2.01µV
=3.01µV(=0.071°C)
Error Calculations with Input Reversal = False
µVError=gainterm+offsetterm
=(830.7µV*0.12%)+(3*0.
67µV+2.0µV)
=0.997µV+4.01µV
=5.01µV(=0.12°C)
Input Error Examples: Type K Thermocouple @ 1300°C
Error in the temperature due to inaccuracy in the measurement of the
thermocouple voltage increases at temperature extremes, particularly when the
temperature and thermocouple type require using the ±200|250 mV range. For
example, assume type K (chromel-alumel) thermocouples are used to measure
temperatures around 1300°C.
These examples demonstrate that at temperature extremes, input offset error is
much less than input gain error because the use of a larger input range is required.
Conditions
CR1000moduletemperature,‐25to50°C
Temperature=1300°C
Referencetemperature=25°C
DeltaT(temperaturedifference)=1275°C
Thermocoupleoutputmultiplierat1300°C=34.9µV°C
‐1
Thermocoupleoutput=1275°C*34.9µV°C
‐1
=44500µV
Inputrange=±250mV
Error Calculations with Input Reversal = True
µVerror=gainterm+offsetterm
=(44500µV*0.12%)+(1.5*66.7µV+1.0µV)
=53.4µV+101.0µV
=154µV(=4.41°C)