Strany 1 - INSTRUCTION MANUAL

INSTRUCTION MANUAL Model 107 Temperature Probe Revision: 3/14 Copyright © 1983- 2014 Campbell Scientific, Inc.

Strany 2

Model 107 Temperature Probe 8. If LoggerNet, PC400, RTDAQ, or PC200W is running on your PC, and the PC to datalogger connection is active, you can cl

Strany 3 - Limited Warranty

Model 107 Temperature Probe Edlog Temp (107) (P11) Polynomial Linearization Error: <±0.1 °C (–24 to 48 °C) <±0.5 °C (–35 to 50 °C) FIGURE 6-

Strany 4 - Assistance

Model 107 Temperature Probe 7.1 Wiring to Datalogger TABLE 7-1. Wire Color, Function, and Datalogger Connection Wire Color Wire Function Datalogger

Strany 5 - Table of Contents

Model 107 Temperature Probe 7.2.1 CRBasic The Therm107() measurement instruction programs CRBasic dataloggers (CR800-series, CR1000, CR3000, CR5000)

Strany 6 - Figures

Model 107 Temperature Probe Standard air temperature measurement heights: • 1.25 to 2.0 m (WMO) • 2.0 m (EPA) • 2.0 m and 10.0 m temperature differ

Strany 7 - Model 107 Temperature Probe

Model 107 Temperature Probe FIGURE 7-2. 107 and 41303-5A Radiation Shield on a CM200 Series Crossarm The 107 is held within the radiation shield by

Strany 8

Model 107 Temperature Probe 8. Operation 8.1 Sensor Schematic FIGURE 8-1. 107 thermistor probe schematic 8.2 Measurement and Output Linearization

Strany 9

Model 107 Temperature Probe TABLE 8-1. 107 Measurement Details Datalogger Model Measurement Instruction Excite mV Result Scaling Equation Applied t

Strany 10 - 6. Specifications

Model 107 Temperature Probe 8.3 Electrically Noisy Environments EMF noise emanating from the ac mains power grid can be a significant source of measu

Strany 11 - 7. Installation

Model 107 Temperature Probe CRBasic For CRBasic dataloggers, the 60 Hz and 50 Hz integration options include a 3 ms settling time; longer settling tim

Strany 13 - 7.2.2 Edlog

Model 107 Temperature Probe Symptom: Temperature is reported as –86 (°C) or –53 (°F). Verify the black wire is connected to the switched excitation c

Strany 14

Appendix A. Importing Short Cut Code This tutorial shows: • How to import a Short Cut program into a program editor for additional refinement. • H

Strany 15

Appendix A. Importing Short Cut Code 6. Import wiring information to the program by opening the associated .DEF file. Copy and paste the section be

Strany 16 - 8. Operation

Appendix B. Example Programs B.1 Example CRBasic Program This example can be used directly with CR800 series, CR1000, CR3000, and CR5000 datalogger

Strany 17

Appendix B. Example Programs ;{CR10X} ;Program measures one 107 temperature probe once a second ;and stores the average temperature every 60 minutes

Strany 18 - 8.4 Long Cable Lengths

Appendix C. Resistance or Voltage Ratio to Temperature TABLE C-1. Voltage Ratio, Resistance, and Temperature1 Actual 100K6A1iA P11 Edlog P11 Edlog

Strany 19 - 9.1 Troubleshooting

Appendix C. Resistance or Voltage Ratio to Temperature 4 283555 14.99 3.99 4.00 5 269034 15.41 5.01 5.00 6 255335 15.83 6.02 6.00 7 242408 16.25 7.03

Strany 21 - A.1.1 CRBasic Datalogger

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbe

Strany 22 - A.1.2 Edlog

Limited Warranty “Products manufactured by CSI are warranted by CSI to be free from defects in materials and workmanship under normal use and service

Strany 23 - B.2 Example Edlog Program

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Strany 24

Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to spec

Strany 25 - Ratio to Temperature

Table of Contents B. Example Programs ... B-1 B.1 Example CRBasic Program ...

Strany 26

Model 107 Temperature Probe 1. Introduction The 107 Temperature Probe uses a thermistor to measure temperature in air, soil, and water. It is compa

Strany 27

Model 107 Temperature Probe 2. The Short Cut installation should place a shortcut icon on the desktop of your computer. To open Short Cut, click on

Strany 28

Model 107 Temperature Probe 5. Under the Available Sensors and Devices list, select the Sensors | Temperature folder. Select 107 Temperature Probe.