Strany 1 - INSTRUCTION MANUAL

INSTRUCTION MANUAL Model 108 Temperature Probe Revision: 3/14 Copyright © 1985- 2014 Campbell Scientific, Inc.

Strany 2

Model 108 Temperature Probe 8. If LoggerNet, PC400, RTDAQ, or PC200W is running on your PC, and the PC to datalogger connection is active, you can cl

Strany 3 - Limited Warranty

Model 108 Temperature Probe Edlog Polynomial (P55) Linearization Error: <±0.5 °C (–5 to 95 °C) <±0.1 °C (–3 to 90 °C) FIGURE 6-1. Lin

Strany 4 - Assistance

Model 108 Temperature Probe 7. Installation If you are programming your datalogger with Short Cut, skip Section 7.1, Wiring to Datalogger, and Sect

Strany 5 - Table of Contents

Model 108 Temperature Probe outlined in Section 8.3, Electrically Noisy Environments, and Section 8.4, Long Cable Lengths. Details of 108 probe measu

Strany 6 - Figures

Model 108 Temperature Probe Variations: • Temperature reported as °F – add a multiplier (P37) of 1.8 and an offset (P34) of 32 • Ac mains noise filt

Strany 7 - Model 108 Temperature Probe

Model 108 Temperature Probe FIGURE 7-1. 108 and 41303-5A Radiation Shield on a tripod mast FIGURE 7-2. 108 and 41303-5A Radiation Shield on a CM20

Strany 8

Model 108 Temperature Probe 7.4 Water Temperature Installation The 108 can be submerged to 15 m (50 ft) or 21 psi. The 108 is not weighted, so a wei

Strany 9

Model 108 Temperature Probe Vs/Vx = 1000 / (Rs + 40000 Ω + 1000 Ω) Solving for Rs: Rs + 41000 Ω = 1000 • (Vx/Vs) Rs = 1000 • (Vx/Vs) – 41000 Ω TABLE 8

Strany 10 - 6. Specifications

Model 108 Temperature Probe C2 = –40.66 C3 = 16.573 C4 = –3.455 C5 = 0.301 1CRBasic dataloggers are CR800, CR1000, CR3000, and CR5000.

Strany 11

Model 108 Temperature Probe 01: Excite-Delay (SE) (P4) 1: 1 Rep 2: 3 ±25 mV slow range ;CR23X,21X,CR7: 50 mV 3: 9 IN Chan 4: 3

Strany 13 - 7.2.2 Edlog

Model 108 Temperature Probe 9.2 Maintenance The 108 probe requires minimal maintenance. For air temperature measurements, check the radiation shield

Strany 14

Appendix A. Importing Short Cut Code This tutorial shows: • How to import a Short Cut program into a program editor for additional refinement. • H

Strany 15

Appendix A. Importing Short Cut Code 6. Import wiring information to the program by opening the associated .DEF file. Copy and paste the section be

Strany 16 - 8. Operation

Appendix B. Example Programs B.1 Example CRBasic Program This example can be used directly with CR800 series, CR1000, CR3000, and CR5000 datalogger

Strany 17

Appendix B. Example Programs ;{CR10X} ;Program measures one 108 temperature probe once a second ;and stores the average temperature every 60 minutes

Strany 18 - 8.4 Long Cable Lengths

Appendix C. Conversion of Thermistor Resistance or Voltage Ratio to Temperature TABLE C-1. Voltage Ratio, Resistance, and Temperature1 Actual Tempe

Strany 19 - 9.1 Troubleshooting

Appendix C. Conversion of Thermistor Resistance or Voltage Ratio to Temperature 31 75913 1.71 30.95 31.00 32 72569 1.76 31.94 32.00 33 69390 1.81 32.

Strany 20 - 9.3 Calibration

Appendix C. Conversion of Thermistor Resistance or Voltage Ratio to Temperature 87 8523 4.04 87.02 87.00 88 8241 4.06 88.01 88.00 89 7970 4.08 88.99

Strany 21 - A.1.1 CRBasic Datalogger

Appendix C. Conversion of Thermistor Resistance or Voltage Ratio to Temperature C-4

Strany 23 - B.2 Example Edlog Program

Limited Warranty “Products manufactured by CSI are warranted by CSI to be free from defects in materials and workmanship under normal use and service

Strany 24

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbe

Strany 25 - Temperature

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Strany 26

Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to spec

Strany 27

Table of Contents B. Example Programs ... B-1 B.1 Example CRBasic Program ...

Strany 28

Model 108 Temperature Probe 1. Introduction The 108 temperature probe uses a thermistor to measure temperature in air, soil, and water. It is compa

Strany 29

Model 108 Temperature Probe 3. When Short Cut opens, select New Program. 4. Select Datalogger Model and Scan Interval (default of 5 seconds is OK

Strany 30

Model 108 Temperature Probe 5. Under the Available Sensors and Devices list, select the Sensors | Temperature folder. Select 108 Temperature Probe.

Komentáře k této Příručce

Žádné komentáře