3WHB10K 3-Wire Half Bridge Terminal Input ModuleRevision: 12/06Copyright © 1996-2006Campbell Scientific, Inc.HLAGHLAGHLG
3WHB10K 3-Wire Half Bridge Terminal Input Module 6 The result of the 3-wire half bridge instruction is equivalent to the ratio of the PRT resistance,
3WHB10K 3-Wire Half Bridge Terminal Input Module 7 To perform the calibration, connect the PRT to the datalogger and program the datalogger to measure
3WHB10K 3-Wire Half Bridge Terminal Input Module 8 This is a blank page.
This is a blank page.
Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com info@campbells
Warranty and Assistance The 3WHB10K 3-WIRE HALF BRIDGE TERMINAL INPUT MODULE is warranted by CAMPBELL SCIENTIFIC, INC. to be free from defects in mate
i 3WHB10K Table of Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat® bookmarks tab
This is a blank page.
1 3WHB10K 3-Wire Half Bridge Terminal Input Module 1. Function Terminal input modules connect directly to the datalogger's input terminals to pr
3WHB10K 3-Wire Half Bridge Terminal Input Module 2 3. Wiring ShieldABHLG10 kΩDataloggerVxHL or AG or G FIGURE 3-1. 3-Wire Half Bridge Used to Measur
3WHB10K 3-Wire Half Bridge Terminal Input Module 3 TABLE 4-1. Excitation Voltage for 100 Ohm PRT in 3WHB10K Based on Maximum Temperature and Input V
3WHB10K 3-Wire Half Bridge Terminal Input Module 4 4.2 21X 1: 3W Half Bridge (P7) 1: 1 Reps 2: 3 ± 50 mV Slow Range 3: 1 SE Channel 4:
3WHB10K 3-Wire Half Bridge Terminal Input Module 5 4.4 CR9000X 'CR9000X Datalogger Public RS_Ro, Temp_F DataTable (Temp_F,1,-1) DataInterval (