A100LK Anemometer7/12Copyright © 2012Campbell Scientific, Inc.
A100LK Anemometer 3. Select A100LK and select the right arrow to add it to the list of sensors to be measured then select next. 4. Select the outp
A100LK Anemometer 5. Wire according to the wiring diagram generated by SCWin Short Cut. 5. Overview The A100LK is an IEC Class 1 anemometer. It i
A100LK Anemometer 6. Specifications Features: • IEC Class 1 performance • Low power • High rate of pulses per revolution (up to 13) produced make
A100LK Anemometer 7. Operation 7.1 Siting Locate wind sensors away from obstructions (e.g., trees and building). As a general rule, there should be
A100LK Anemometer 7.3 Programming with CRBasic and Edlog Wind speed is typically measured using a pulse channel. For dataloggers using CRBasic, use
A100LK Anemometer 'Main Program BeginProg Scan (1,Sec,0,0) PanelTemp (PTemp,250) Battery (batt_volt) 'A100LK Wind Speed Sens
A100LK Anemometer 10 10. References 1. IEC 61400 Part 12-1, “Wind turbine generator systems Part 12: Wind Turbine Power Performance Testing”. 2. ht
Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel
Warranty “PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materi
Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing
A100LK Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links t
A100LK Anemometer 1. Introduction The Windspeed Ltd Model A100LK Anemometer is an IEC and MEASNET rated instrument recommended for wind energy applic
A100LK Anemometer 4. Quickstart 4.1 Step 1 — Mount the Sensor To mount the sensor, refer to Figure 4-1 and do the following. 1. Mount a CM202, CM20
A100LK Anemometer 4.2 Step 2 — Use SCWin ShortCut to Program Datalogger and Generate Wiring Diagram The simplest method for programming the datalogge