Section 8. Operation
300
Figure 94: Pulse-input channels
8.1.5.1.1 High-frequency Pulse (P1 - P2)
High-frequency pulse inputs are routed to an inverting CMOS input buffer with
input hysteresis. The CMOS input buffer is an output zero level with its input
2.2 V, and an output one level with its input 0.9 V. When a pulse channel is
configured for high-frequency pulse, an internal 100-kΩ pull-up resistor to 5 Vdc
on the P1 or P2 input is automatically employed. This pull-up resistor
accommodates open-collector (open-drain) output devices for high-frequency
input.
8.1.5.1.2 Low-Level ac (P1 - P2)
Rotating magnetic-pickup sensors commonly generate ac output voltages ranging
from thousandths of Volts at low rotational speeds to several volts at high
rotational speeds. Pulse channels contain internal signal-conditioning hardware
for measuring low-level ac-output sensors. When configured for low-level ac, P1
through P2 measure signals ranging from 20-mV RMS (±28 mV peak) to 14-V
RMS (±20 V peak). Internal ac coupling is incorporated in the low-level ac
hardware to eliminate dc offset voltages of up to ±0.5 Vdc.
8.1.5.1.3 Switch Closure (P1 - P2)
Switch-closure mode measures switch closure events, such as occur with a
common tipping bucket rain gage. An internal 100-kΩ pull-up resistor pulls an
input to 5 Vdc with the switch open, whereas a switch closure to ground pulls the
input to 0 V. An internal 3.3-ms time-constant RC-debounce filter eliminates
multiple counts from a single switch closure event.