Campbell-scientific CR800 and CR850 Measurement and Control Systems Uživatelský manuál Strana 166

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 566
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 165
Section 7. Installation
166
Figure 50: Strain-gage shunt calibration started
Figure 51: Strain-gage shunt calibration finished
7.8.1.6.2 Quarter-Bridge Zero (Option 10)
Continuing from Quarter-Bridge Shunt (Option 13)
(p. 165), keep the 249-k
resistor in place to simulate a strain. Using the external keyboard / display or
software numeric monitor, change the value in variable Zero_Mode to 1 to start
the zero calibration as shown in figure Starting Zero Procedure (p. 166). When
Z
ero_Mode increments to 6, zero calibration is complete as shown in figure Zero
Procedure Finished
(p. 167).
Figure 52: Starting zero procedure
Zobrazit stránku 165
1 2 ... 161 162 163 164 165 166 167 168 169 170 171 ... 565 566

Komentáře k této Příručce

Žádné komentáře