APPENDIX G. DATALOGGER INITIATED COMMUNICATIONS
G-2
6: Do (P86)
1: 51 Set Port 1 Low
7: If time is (P92)
1: 0 Minutes (Seconds --) into a
2: 60 Interval (same units as above)
3: 10 Set Output Flag High
8: Real Time (P77)
1: 120 Day,Hour/Minute (2400 at
midnight)
9: Average (P71)
1: 2 Reps
2: 1 Loc [ AIR_TEMP ]
10: Maximize (P73)
1: 2 Reps
2: 0 Value Only
3: 1 Loc [ AIR_TEMP ]
11: Sample (P70)
1: 2 Reps
2: 3 Loc [ BATT_VOLT ]
12: IF (X<=>F) (P89)
1: 1 X Loc [ AIR_TEMP ]
2: 4 <
3: 31 F
4: 15 Set Flag 5 High
13: IF (X<=>F) (P89)
1: 1 X Loc [ AIR_TEMP ]
2: 3 >=
3: 32 F
4: 25 Set Flag 5 Low
14: Initiate Telecommunications (P97)
1: 21 Phone Modem/1200 Baud
2: 5 Disabled when User Flag 5
is High
3: 75 Seconds Call Time Limit
4: 120 Seconds Before Fast Retry
5: 2 Fast Retries
6: 5 Minutes before Slow Retry
7: 4 Failures Loc [ Failures ]
8: 115 Data Logger ID
15: Extended Parameters (P63)
1: 1 Option
2: 4 Option
3: 3 Option
4: 5 Option
5: 7 Option
6: 5 Option
7: 2 Option
8: 3 Option
16: Extended Parameters (P63)
1: 2 Option
2: 6 Option
3: 8 Option
4: 13 Option
5: 00 Option
6: 00 Option
7: 00 Option
8: 00 Option
G.3 PC208W WINDOWS SOFTWARE
AND COMPUTER SETUP
Using Campbell Scientific’s PC208W Windows
software for callback, the network or Device
Map must be setup for the site as if the
computer is initiating the call. Two additional
settings are required to allow the datalogger to
initiate the call.
To have PC208W answer a datalogger initiated
call, you must first Setup the Connection to the
site. Figure G.3-1 shows an example station for
a CR10X site called “Logan”. Below is an
example of the three devices required for the
phone link to the datalogger with the key
settings required to allow callback.
COM Port: (Figure G.3-2)
Correct COM port number (COM2
for this example)
COM port Hardware settings has
“Allow Callback on This Port”
Selected
Modem (Modem at computer): (Figure G.3-3)
Correct baud rate: 1200 for DC112
or VS1, 9600 for COM200/COM300.
(9600 for this example)
Correct modem type and dialing
strings. If there is an “&Q#” in the
dialing string, it must be an “&Q5”
for 9600 Baud or “&Q0” for 1200
Baud. (&Q5 for this example)
Datalogger: (Figure G.3-1)
Correct Datalogger type (CR10X for
this example)
On the Hardware screen, the
Callback ID# must be the 3 Digit
ID# used in your program. (115 for
this example)