CS135 Ceilometer
26
Line Example line output
15 HOffset A
Description of the line sections
Section Description
A Height offset reported in feet or metres dependent upon the UNITS command
Line Exam
p
le line out
p
ut
16 LaserRunDays A
Descri
p
tion of the line sections
Section Description
A Number of days that the laser module has been active for
Line Example line output
17 Flags 0000 0000 0000
Description of the most significant alarm word (left side, bits going left to right). Each alarm word is
a hexadecimal sum of all the error bits.
Bit Description
8000 XXXX XXXX Units. Feet = 0, metre = 1.
4000 XXXX XXXX Reserved for future use
2000 XXXX XXXX Reserved for future use
1000 XXXX XXXX Reserved for future use
0800 XXXX XXXX Reserved for future use
0400 XXXX XXXX Laser shut down due to operating temperature out of range
0200 XXXX XXXX The lead acid battery voltage is reading low
0100 XXXX XXXX Mains supply has failed (Required a PSU to be present)
0080 XXXX XXXX The external heater blower assembly temperature is out of bounds
0040 XXXX XXXX External heater blower failure
0020 XXXX XXXX The PSUs internal temperature is high
0010 XXXX XXXX PSU OS has failed its signature check
0008 XXXX XXXX No communications between DSP and PSU
0004 XXXX XXXX Photo diode and Laser windows are dirty. This can only be set if the laser is on
0002 XXXX XXXX Tilt beyond limit set by user, default 45 degrees
0001 XXXX XXXX No communications between DSP and inclinometer board
Description of the middle alarm word (middle word, bits going left to right)
Bit Description
XXXX 8000 XXXX The sensors internal humidity is high
XXXX 4000 XXXX Communications to the DSP boards temperature and humidity chip have failed
XXXX 2000 XXXX DSP input supply voltage is low
XXXX 1000 XXXX Self-test active
XXXX 0800 XXXX Watch dog counter updated
XXXX 0400 XXXX User setting stored in flash failed their signature checks
XXXX 0200 XXXX DSP factory calibration stored in flash has failed its signature check
XXXX 0100 XXXX DSP board OS signature test failed
XXXX 0080 XXXX DSP board RAM test failed
XXXX 0040 XXXX DSP boards on board PSUs are out of bounds
XXXX 0020 XXXX TOP board non-volatile storage is corrupt
XXXX 0010 XXXX TOP board OS signature test has failed
XXXX 0008 XXXX TOP boards ADC and DAC are not within specifications
XXXX 0004 XXXX TOP boards on board PSUs are out of bounds
XXXX 0002 XXXX Communications have failed between TOP board and the DSP
XXXX 0001 XXXX Photo diode background radiance is out of range
Description of the least significant alarm word (right side, bits going left to right)
Bit Description
XXXX XXXX 8000 Photo diode temperature is out of range
XXXX XXXX 4000 Photo diode is saturated
XXXX XXXX 2000 Photo diode calibrator temperature is out of range
Zobrazit stránku 33