Strany 1 - INSTRUCTION MANUAL

INSTRUCTION MANUAL CSIM11 pH and ORP Probes Revision: 7/14 Copyright © 2000- 2014 Campbell Scientific, Inc.

Strany 2

CSIM11 pH and ORP Probes 2. The Short Cut installation should place a Short Cut icon on the desktop of your computer. To open Short Cut, click on th

Strany 3 - Limited Warranty

CSIM11 pH and ORP Probes 5. Under the Available Sensors and Devices list, select the Sensors | Temperature folder. Select 107 Temperature Probe. Cl

Strany 4 - Assistance

CSIM11 pH and ORP Probes 7. After selecting the sensors, click at the left of the screen on Wiring Diagram to see how the sensor is to be wired to th

Strany 5 - Precautions

CSIM11 pH and ORP Probes 2. The Short Cut installation should place a Short Cut icon on the desktop of your computer. To open Short Cut, click on th

Strany 6

CSIM11 pH and ORP Probes 5. Under the Available Sensors and Devices list, select the Sensors | Water | Quality folder. Select CSIM11 ORP Probe. Cli

Strany 7 - Table of Contents

CSIM11 pH and ORP Probes 7. Select any other sensors you have, then finish the remaining Short Cut steps to complete the program. The remaining step

Strany 8

CSIM11 pH and ORP Probes 6. Specifications Features: • Internal amplifier boosts the signal, decreasing signal interference • Titanium ground rod

Strany 9

CSIM11 pH and ORP Probes Drift: < 2 mV per week Power: Two 3 Vdc lithium batteries that should last the lifetime of the sensor Length: 17.8 cm (

Strany 10

CSIM11 pH and ORP Probes 7.3 Wiring The CSIM11 and CSIM11-ORP are connected to differential channels. Connections to Campbell Scientific dataloggers

Strany 11

CSIM11 pH and ORP Probes Short Cut cannot edit programs after they are imported and edited in CRBasic Editor. A Short Cut tutorial is available in Sec

Strany 13

CSIM11 pH and ORP Probes 8. Calibration/ORP Check 8.1 pH Sensor Calibration Calibration should be carried out according to the detailed procedure la

Strany 14

CSIM11 pH and ORP Probes • Coatings can cause 1) Slower Response; 2) Zero Offset increase; 3) Dehydration • Steam Sterilization causes 1) Dehydrati

Strany 15 - 5. Overview

CSIM11 pH and ORP Probes 10 minutes each. After treating the electrode with these strong acids or bases, rinse the electrode with water and soak it i

Strany 16 - 6. Specifications

CSIM11 pH and ORP Probes sensor in 3.5 molar KCl, or water if KCl is not available, and heat to approximately 60 °C for 15 minutes; retest. SYMPTOM:

Strany 17 - 7. Installation

CSIM11 pH and ORP Probes 16

Strany 18 - 7.4 Datalogger Programming

Appendix A. Importing Short Cut Code This tutorial shows: • How to import a Short Cut program into a program editor for additional refinement. • H

Strany 19

Appendix A. Importing Short Cut Code 6. Import wiring information to the program by opening the associated .DEF file. Copy and paste the section be

Strany 20 - 9. Maintenance

Appendix B. Example Programs B.1 CRBasic Program The following CR1000 program measures a CSIM11 and 107 temperature probe and has the datalogger pr

Strany 21 - 9.2 Electrode Cleaning

Appendix B. Example Programs B.2 Edlog Programs The following CR10X programs measure a CSIM11 and 107 temperature probe and have the datalogger pro

Strany 22 - 10. Troubleshooting

Appendix B. Example Programs *Table 2 Program 02: 0.0000 Execution Interval (seconds) *Table 3 Subroutines End Program B.2.2 CR10(X) Exampl

Strany 23

Limited Warranty “Products manufactured by CSI are warranted by CSI to be free from defects in materials and workmanship under normal use and service

Strany 24 - CSIM11 pH and ORP Probes

Appendix B. Example Programs 9: Z=X*Y (P36) 1: 2 X Loc [ pH ] 2: 3 Y Loc [ pHMult ] 3: 2 Z Loc [ pH ] ;The follo

Strany 25 - A.1.1 CRBasic Datalogger

Appendix C. Detailed Calibration Procedure and Manufacturer Tips C.1 Calibration Good laboratory practices should be used and protective gloves an

Strany 26 - A.1.2 Edlog

Appendix C. Detailed Calibration Procedure and Manufacturer Tips 7. Remove and rinse the electrode, then place it in the 10.00 buffer. 8. In the da

Strany 28 - B.2 Edlog Programs

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbe

Strany 29 - Expression Editor

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Strany 30

Precautions DANGER — MANY HAZARDS ARE ASSOCIATED WITH INSTALLING, USING, MAINTAINING, AND WORKING ON OR AROUND TRIPODS, TOWERS, AND ANY ATTACHMENTS TO

Strany 32 - C.2 Tips and Techniques

Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to spec

Strany 33

Table of Contents B. Example Programs .... B-1 B.1 CRBasic Program ...

Strany 34

CSIM11 pH and ORP Probes 1. Introduction The CSIM11 measures the full pH range of liquids, and the CSIM11-ORP measures oxidation reduction potential