CS650 and CS655 Water Content Reflectometers
CS650 CS655
Volumetric Water
Content using Topp
Equation (m
3
/m
3
)
Range:
5% to 50% 5% to 50%
Accuracy†:
±3% VWC typical in
mineral soils where
solution EC 3 dS/m
±3% VWC typical in
mineral soils where
solution EC 10 dS/m
Precision‡:
<0.05% <0.05%
Electrical
Conductivity
Range Solution
EC:
0 to 3 dS/m
0 to 8 dS/m
Range Bulk EC:
0 to 3 dS/m 0 to 8 dS/m
Accuracy†:
±(5% of reading +
0.05 dS/m)
±(5% of reading +
0.05 dS/m)
Precision‡:
0.5% of BEC 0.5% of BEC
Temperature
Soil
Measurement
Range:
-10 to + 70°C
-10 to + 70°C
Operational
Range:
0 to + 70°C
0 to + 70°C
Accuracy†:
±0.5°C for probe
body buried in soil
±0.5°C for probe
body buried in soil
Precision‡:
±0.02°C ±0.02°C
Sensing Volume*:
7800 cm
3
3600 cm
3
*Sensing Volume approximately 7.5 cm radius around each probe rod
and 4.5 cm beyond the end of the rods
Accuracy specifications are based on laboratory measurements in a
series of solutions with dielectric permittivities ranging from 1 to 81
and solution electrical conductivities ranging from 0 to 3 dS/m.
Precision describes the repeatability of a measurement. It is
determined for the CS650 by taking repeated measurements in the
same material. The precision of the CS650 is better than 0.05 %
volumetric water content and 0.01 dS/m electrical conductivity.