CS547 Conductivity and Temperature Probe and A547 Interface
6
9: Full Bridge (P6)
1: 1 Reps
2: 15 ±2500 mV Fast Range
3: 1 DIFF Channel
4: 1 Excite all reps w/Exchan 1
5: 2500 mV Excitation
6: 1 Loc [ Rs ]
7: -.001 Mult
8: 1 Offset
10: BR Transform Rf[X/(1-X)] (P59)
1: 1 Reps
2: 1 Loc [ Rs ]
3: 1 Multiplier (Rf)
11: End (P95)
12: If Case Location < F (P83)
1: 280 F
2: 30 Then Do
13: Full Bridge (P6)
1: 1 Reps
2: 14 ±250 mV Fast Range
3: 1 DIFF Channel
4: 1 Excite all reps w/Exchan 1
5: 2500 mV Excitation
6: 1 Loc [ Rs ]
7: -.001 Mult
8: 1 Offset
14: BR Transform Rf[X/(1-X)] (P59)
1: 1 Reps
2: 1 Loc [ Rs ]
3: 1 Multiplier (Rf)
15: End (P95)
16: End (P95)
;
;Subtract resistance errors (Rp) caused by the blocking capacitors
;(0.005Kohm) and the cable length (0.000032kohm/ft). Enter cable lead
;length in nnn below.
;
17: Z=F (P30)
1: nnn F Enter cable length in feet.
2: 00 Exponent of 10
3: 5 Z Loc [ Rp ]
18: Z=X*F (P37)
1: 5 X Loc [ Rp ]
2: .00032 F
3: 5 Z Loc [ Rp ]
(cont.)