CS470/CS471 Compact Bubbler System
5. Under the Available Sensors and Devices list, select the Sensors |
Generic Measurements folder. Select SDI-12 Sensor, click to move
the selection to the Selected device window. Enter the result names and
units. The default SDI-12 address is 0 and the default SDI-12 command is
aM!. The SDI-12 address should only be changed from the default when
more than one sensor is connected to the same datalogger channel. The
SDI-12 command can be changed by clicking the SDI-12 Command box
and selecting one of the other options.
6. After selecting the sensor, click at the left of the screen on Wiring
Diagram to see how the sensor is to be wired to the datalogger. The
wiring diagram can be printed out now or after more sensors are added.