CS450/CS455 Pressure Transducer
CR10X-PB and CR510-PB dataloggers require “*#n” being entered at the
prompt, where n is the control port being used.
Now check for response from the sensor with address zero by typing the SDI-
12 Identify command “0I!<Enter>” (that’s a zero, not the letter O). The sensor
should respond with an identification string similar to
“013CSI450.Std.01_xxxxxxxx”, where xxxxxxxx represents the eight-digit
serial number. Note that the SDI-12 standard allows for multiple probes to be
connected to one datalogger control port. For example, if you have another
SDI-12 probe on the C1 that has address 7, you could issue the identify
command “7I!<Enter>”.
Only one sensor of the same address can be connected when using the change-
address command.
3.1.4 RS-232 Connection via PC Terminal Software
PC terminal software can be used to communicate with the CS450/CS455 via
the RS-232 communication mode. For this example, HyperTerminal will be
used (see FIGURE 3-6).
The CS450/CS455 is connected to the PC via the A200 (see Section 3.1.1,
Using the A200).
Upon setup, HyperTerminal will request you enter the Communication
connection; it defaults to a phone connection. Change the communication to
the appropriate “Com” in the “Connect Using” box (see Section 3.1.1.4,
Determining which COM Port the A200 has been Assigned, to determine the
COM port that was assigned to the A200). HyperTerminal will then prompt
for the proper “Port Settings”. These should be:
TABLE 3-3. RS-232 Settings
Bits per second 9600
Data bits 8
Parity None
Stop bits 1
Flow control None
From the HyperTerminal screen, you will now be able to communicate with
the CS450/CS455. At the prompt, push the Enter key several times. This will
wake-up the RS-232 mode of the sensor. By default, the CS450/CS455 is in
the SDI-12 mode for communication. Once in the RS-232 mode, if there is no
communication for 20 seconds, the sensor will return to the SDI-12 mode.
9
Zobrazit stránku 14
1 2 ... 10 11 12 13 14 15 16 17 18 19 20 ... 33 34

Komentáře k této Příručce

Žádné komentáře

Muvit MUSLI0537 manuals

Owner’s manuals and user’s guides for Mobile phone cases Muvit MUSLI0537.
We providing 1 pdf manuals Muvit MUSLI0537 for download free by document types: Datasheet


Muvit MUSLI0537 Datasheet (1 pages)


Brand: Muvit | Category: Mobile phone cases | Size: 0.06 MB |

 

Table of contents





More products and manuals for Mobile phone cases Muvit

Models Document Type
MUCUN0280 Datasheet   Muvit MUCUN0280 mobile phone case, 1 pages
MUSLI0525 Datasheet   Muvit MUSLI0525 mobile phone case, 1 pages
MUCUN0284 Datasheet   Muvit MUCUN0284 mobile phone case, 1 pages
MUSLI0606 Datasheet   Muvit MUSLI0606 mobile phone case, 1 pages
SEBMC0039 Datasheet   Muvit SEBMC0039 mobile phone case, 1 pages
MUCUN0285 Datasheet   Muvit MUCUN0285 mobile phone case, 1 pages
MUSLI0465 Datasheet   Muvit MUSLI0465 mobile phone case, 1 pages
SESLI0086 Datasheet   Muvit SESLI0086 mobile phone case, 1 pages
MUCUN0281 Datasheet   Muvit MUCUN0281 mobile phone case, 1 pages
MUMAG0006 Datasheet   Muvit MUMAG0006 mobile phone case, 1 pages
MUCUN0232 Datasheet   Muvit MUCUN0232 mobile phone case, 1 pages
MUCUN0278 Datasheet   Muvit MUCUN0278 mobile phone case, 1 pages
MUSNS0072 Datasheet   Muvit MUSNS0072 mobile phone case, 1 pages
MUSNS0003 Datasheet   Muvit MUSNS0003 mobile phone case, 1 pages
MUCUN0227 Datasheet   Muvit MUCUN0227 mobile phone case, 1 pages
MUCUN0229 Datasheet   Muvit MUCUN0229 mobile phone case, 1 pages
MUSLI0542 Datasheet   Muvit MUSLI0542 mobile phone case, 1 pages
MUSLI0563 Datasheet   Muvit MUSLI0563 mobile phone case, 1 pages
MUSLI0303 Datasheet   Muvit MUSLI0303 mobile phone case, 1 pages
SEMAG0002 Datasheet   Muvit SEMAG0002 mobile phone case, 1 pages